logo
Good price  online

products details

Created with Pixso. Home Created with Pixso. Products Created with Pixso.
Multiparameter Water Quality Monitoring
Created with Pixso.

Digital Full Spectrum Multi Parameter Water Quality Sensor TOC BOD UV254 Color Sensor

Digital Full Spectrum Multi Parameter Water Quality Sensor TOC BOD UV254 Color Sensor

Brand Name: JUGE
Model Number: AG600
MOQ: 1 Unit
Price: USD1000-5000
Packaging Details: Neutral carton
Payment Terms: Online order, T/T
Detail Information
Place of Origin:
Nanjing, China
Certification:
IOS9001, CE
Name:
Full Spectrum Sensor
Parameter:
COD、 Nitrate Nitrogen, Turbidity,TOC、BOD、UV254、 Color And Temperature (OEM TSS, Nitrite Nitrogen, Etc.)
Principle:
Absorption Method
Detection Wavelength:
190-760 Nm
Protection Level:
IP68、under Water 10m
Material:
S316L
Communication:
RS485(Modbus RTU)、Maximum Baud Rate 115200 Bps
Power:
(12~28)V DC /2W(In Non Cleaning Mode)
Size/weight:
2500g/φ50mm ×L400mm
Supply Ability:
100 Unit/Week
Highlight:

Full Spectrum Water Quality Sensor

,

Multi Parameter Water Quality Sensor

,

Multi Parameter online water monitoring system

Product Description

Digital Full Spectrum Water Quality Multi Parameter Sensor COD Nitrate Nitrogen Turbidity TOC BOD UV254  Color Sensor

 

>>Overview

The AG-600 digital full-spectrum multi-parameter sensor operates based on the Lambert-Beer absorption law. It employs a miniature xenon flash lamp to irradiate water samples, where the absorbed light intensity correlates with measured components. Designed with digital and intelligent sensor principles, it automatically compensates for measurement deviations caused by light source fluctuations, component aging, and temperature variations. Featuring standardized digital signal output capability, this sensor enables direct networking and system integration without requiring additional controllers.

 

>>Feature

● Detection wavelength: 190 nm to 760 nm

● Corrosion-resistant housing: S316L stainless steel material

● No chemical reagents required, zero secondary pollution, rapid measurement

● Air purge or automatic electric brush cleaning: Prevents measurement light path blockage by deposits, enabling maintenance-free operation

● Long-life xenon flash lamp: Service life exceeding 10 years

● Built-in light intensity reference detection: Ensures excellent data stability

● Data analysis software: Features standardization, logging, analysis, and diagnostic capabilities

● RS485 output: Strong anti-noise and anti-interference performance, delivering enhanced data stability

● Calibration log storage: Plug-and-play design; no initial calibration required upon first use

● User-friendly operation: Supports online software upgrades

 

>>Technical Parameter

Model

AG-600

Parameter

COD, Nitrate nitrogen, Turbidity,TOC,BOD,UV254, Color and temperature

(OEM TSS, Nitrite Nitrogen, etc.)

Principle

Absorption method

 

 

 

 

Range

Path

2mm

10mm

20mm

COD(mg/L)

0~2000

0~400

0~200

Nitrate nitrogen(mg/L)

0 ~50

0 ~10

0 ~5

turbidity(NTU)

0~800

0~200

0 ~100

SAC254(abs/m)

0~750

0 ~150

0~75

TOC(mg/L)

0~800

0 ~160

0~80

BOD(mg/L)

0 ~1500

0 ~300

0 ~150

Color(Hazen)

0 ~5000

0 ~1000

0 ~500

Temperature(℃)

0~50

Accuracy

COD,TOC, BOD,Nitrate nitrogen:≤±3%;Turbidity,Color:≤±5%

Repeatability

COD,TOC, BOD,Nitrate nitrogen:≤2%;Turbidity,Color:≤3%

Temperature Range

(0~50)℃

Protection level

IP68,under water 10m

Cleaning

Mechanical brush automatic cleaning+air blowing

Communication

RS485(Modbus RTU),Maximum baud rate 115200 bps

Power

(12~28)V DC /2W(In non cleaning mode)

Material

S316L

Size/weight

2500g/φ50mm ×L400mm