Good price  online

products details

Created with Pixso. Home Created with Pixso. Products Created with Pixso.
Online Water Quality Analyzer
Created with Pixso.

Industrial Sewage Online Ammonia nitrogen Water Quality Analyzer

Industrial Sewage Online Ammonia nitrogen Water Quality Analyzer

Brand Name: JUGE
Model Number: AG9001
MOQ: 1 Unit
Price: USD2000-7000
Packaging Details: Neutral carton
Payment Terms: Online order,T/T
Detail Information
Place of Origin:
Nanjing, China
Certification:
IOS9001, CE
‌Test Method‌:
Salicylic Acid Spectrophotometric Method
‌Measuring Range‌:
0~300mg/L (segmented Measurement, Expandable)
‌Indication Error‌:
≤±5%, ≤±8%, ≤±10%
‌Quantitative Lower Limit‌:
≤0.15mg/L
‌Repeatability‌:
≤2%
‌Power Supply‌:
AC 220±10% V
‌Dimensions‌:
520H × 370W × 265D (mm)
Weight‌:
40kg
Supply Ability:
100 Unit/Week
Highlight:

300mg/l water quality monitor

,

300mg/l water quality analyzer

,

spectrophotometric method water quality monitor

Product Description

Industrial Sewage Online Ammonia nitrogen Water Quality Analyzer


>>Overview

Ammonia nitrogen in water refers to ammonia existing in its free form, primarily originating from the decomposition of nitrogenous organic matter in domestic wastewater by microbial activity, industrial effluents such as coking and synthetic ammonia production, and agricultural drainage. Elevated ammonia nitrogen levels in water bodies pose toxic effects on aquatic life and varying degrees of harm to humans. Measuring ammonia nitrogen concentrations aids in assessing water pollution levels and "self-purification" status, making it a critical indicator of water quality contamination.

This analyzer is designed for ‌long-term unattended and automatic continuous operation‌ based on field configurations. It is widely applicable to scenarios including industrial wastewater discharge, municipal sewage treatment plant effluent, and environmental quality surface water monitoring. Depending on the complexity of on-site conditions, optional preprocessing systems can be integrated to ensure ‌reliable testing processes‌, ‌accurate results‌, and full compliance with diverse on-site requirements.


>>Principle
The analyzer employs the ‌salicylate colorimetric method‌. After mixing the water sample with a masking agent, ammonia nitrogen (existing as free ammonia or ammonium ions) reacts with salicylate ions and hypochlorite ions under alkaline conditions in the presence of a sensitizing agent, forming a colored complex. The analyzer detects this chromatic change, converts it into an ammonia nitrogen concentration value, and outputs the result. ‌The amount of colored complex formed corresponds to the ammonia nitrogen concentration.


>>Technical Parameter

Test Method

Salicylic Acid Spectrophotometric Method

Measuring Range

0~300mg/L (segmented measurement, expandable)

Indication Error

20% range standard solution: ≤±10%

50% range standard solution: ≤±8%

80% range standard solution: ≤±5%

Quantitative Lower Limit

≤0.15mg/L

Repeatability

≤2%

Drift (24h)

Low concentration drift: ≤0.02mg/L

High concentration drift:  ≤1%

Measurement Time

<30min (digestion time adjustable)

Measurement Modes

Adjustable time interval, fixed time point, or triggered measurement

Calibration Modes

Automatic calibration (1~99 days adjustable)

Manual calibration based on actual water samples

Maintenance Cycle

1 month per session; ~30min per maintenance

Human-Machine Interface

Touchscreen display and command input

Self-Diagnosis & Protection

Self-diagnoses operational status; retains data during anomalies/power outages

Automatically clears residual reactants and resumes operation after reset/power recovery

Data Storage

5-year data capacity

Input Interface

Digital input

Output Interfaces

1× RS232, 1× RS485

2× 4~20mA analog outputs

Operating Environment

Indoor use; recommended temperature: 5~28℃, humidity ≤90% (non-condensing)

Power Supply

Voltage: AC 220±10% V

Frequency: 50±0.5 Hz

Power: ≤150W (excluding sampling pump)

Dimensions

520H × 370W × 265D (mm)